The DHS is a data logging system for validating dry heat sterilization cycles. Includes standard calibration certificate.
Key Features:
Specification:
|
GENERAL |
|
|
Material |
316 Stainless Steel, PTFE, Silicon |
|
Reading Rate |
4 readings per second up to 1 reading every 24 hours |
|
Battery Life |
1 year typical (1 minute reading rate at +25 °C / +77 °F) User Replaceable Battery |
|
Dimensions |
1.75 in OD x 9.6 in L (44.45 mm OD x 243.8 mm L) |
|
Operating Environment |
-200 °C to 400 °C 0 %RH to 95 %RH non-condensing |
|
Required Interface Package |
IFC400 or IFC406 |
| Approvals | CE |
|
TEMPERATURE SENSOR |
|
|
Temperature Sensor |
36 inch Glass Braided RTD depyrogenation probe |
|
Temperature Range |
-200 °C to +400 °C |
| Temperature Resolution |
0.0001 Ω, 0.01 °C (0.02 °F) |
| Temperature Calibrated Accuracy |
±1.2 °C (-200 °C to +400 °C) |